Patient Monitor User Manual
Physiological alarms:
Message
HR HIGH
HR LOW
Technical alarms:
Message
ECG LEAD OFF
ECG LL LEAD
OFF
ECG LA LEAD
OFF
ECG RA LEAD
OFF
ECG
V
LEAD
OFF
ECG
SIGNAL
EXCEED
ECG
COMM
STOP
HR measuring value is above the upper alarm limit. User-selectable
HR measuring value is below the lower alarm
limit.
Cause
1) The drive lead or more
than
one
ECG
electrode falls off the skin;
2) ECG cables fall off the
monitor.
ECG electrode LL falls off
the skin or the ECG cable
LL falls off the monitor.
ECG electrode LA falls off
the skin or the ECG cable
LA falls off the monitor.
ECG electrode RA falls off
the skin or the ECG cable
RA falls off the monitor.
ECG electrode V falls off
the skin or the ECG cable
V falls off the monitor.
ECG measuring value is
beyond measuring range.
ECG module failure or
communication failure
Cause
Alarm level
limb
Low
Low
Low
Low
Low
High
High
- 87 -
ECG/ RESP Monitoring
Alarm level
User-selectable
What to do
Make
sure
that
electrodes,
leads
patient cables are properly
connected.
Make
sure
that
electrodes,
leads
patient cables are properly
connected.
Make
sure
that
electrodes,
leads
patient cables are properly
connected.
Make
sure
that
electrodes,
leads
patient cables are properly
connected.
Check lead connection
and patient condition
Stop
using
measuring
function of ECG module,
notify
biomedical
engineer
manufacturer's
service
staff.
all
and
all
and
all
and
all
and
or