Chapter 8 Recording (Optional)

General information on recording

Instructions for configuring and recording
Recording messages
8.1 General Information on Recording
A thermal dot matrices recorder with 48mm wide printout paper is used for the monitor.
Performance of the Recorder
Waveform record is printed out at the rate of 25 mm/s or 50 mm/s.
It can record up to three waveforms.
English printout.
User-selectable real-time recording time and waveform.
Auto recording interval is set by the user, and the waveform is in accordance with the real
time recording.
NOTE:
It is suggested that the user should not use the recorder when the low battery displays, or
the monitor may be turned off automatically.

8.2 Recording Type

The monitor provides several types of stripe recording:
Continuous real-time recording
8 second real-time recording
Automatic interval recording
Physiological alarm recording
Frozen waveform recording
Trend graph/table review recording
NIBP review recording
Alarm event review recording
Arrhythmia review recording
Titration table recording
NOTE:
1
When ECG waveforms are selected for printing, with gain of ×1, ×0.5 or ×0.25, a
3-channel waveform can be printed out; however, with gain of ×2, only a 2-channel
waveform can be printed out to avoid overlapping of waveforms, and the third
waveform will be omitted.
2
The 3-channel waveform can be printed only in real-time recording, while it is not
- 56 -
Recording (Optional)