alarm function, and prompt the
In order to avoid endangering the patient's life, the user should use this function
cautiously.
ALM LEV: used to set up the alarm level, selectable from HIGH, MED or LOW.
ALM REC: used to start/stop recording TEMP alarms. Pick ON to enable report printing
upon TEMP alarm.
Alarm for T1, T2, TD occurs when the measured temperature exceeds the set alarm upper
limit or falls below alarm lower limit.
T1 represents the TEMP of Channel 1, T2 represents the TEMP of Channel 2, TD represents the
TEMP difference of T1 and T2.
UNIT: To set temperature unit (ºC or ºF).
DEFAULT: Pick this item to access the TEMP DEFAULT CONFIG dialog box, in which
the user may select whether the FACTORY DEFAULT CONFIG or the USER
DEFAULT CONFIG is to be used. After selecting any of the items and exiting the dialog
box, the system will pop up the dialog box asking for the user's confirmation.

15.3 TEMP Alarm Message

Tables below describe the possible physiological alarms, technical alarms occurring during
TEMP measurement.
Physiological alarms:
Message
T1 HIGH
T1 LOW
T2 HIGH
T2 LOW
TD HIGH
symbol besides TEMP numeric.
WARNING
Cause
Measuring value of T1 channel is
above upper alarm limit.
Measuring value of T1 channel is
below lower alarm limit.
Measuring value of T2 channel is
above upper alarm limit.
Measuring value of T2 channel is
below lower alarm limit.
Temperature difference of T1 and T2
is
above
upper
difference limit.
- 120 -
User-selectable
User-selectable
User-selectable
User-selectable
temperature
User-selectable
TEMP Monitoring
Alarm Level