Technical alarms: (display in the area below the NIBP value)
Message
NIBP
COMM
STOP
NIBP AIR LEAK
NIBP
OVER
PRESSURE
NIBP
INIT
PRESSURE
HIGH
NIBP
SECONDARY
OVER
PRESSURE
NIBP TIME OUT
NIBP Self Test
Error
NIBP
CUFF
TYPE ERROR
Cause
NIBP module failure or
communication failure
NIBP pump, valve, cuff or
tube has a leakage.
Pressure has exceeded the
specified
upper
limit.
The initial pressure is too
high during measuring
Secondary over pressure
protection
Measuring
time
exceeded
120
seconds
(adult) or 90 seconds
(neonatal).
When
the
monitor
powered
on,
module is detected to fail
in calibration.
The cuff type used isn't
consistent with the patient
type.
Alarm
Level
Stop using measuring function
of
High
biomedical
Manufacturer's service staff.
Check and replace the leaking
parts,
Low
biomedical
manufacturer's service staff.
Measure
persists, stop using measuring
safety
Low
function of NIBP module and
notify biomedical engineer or
manufacturer's service staff.
Measure
persists, stop using measuring
Low
function of NIBP module and
notify biomedical engineer or
manufacturer's service staff.
Notify biomedical engineer or
High
manufacturer's service staff.
has
Measure again or use other
Low
measuring method.
is
NIBP
Low
Contact your service personnel.
Confirm the patient type and
Low
change the cuff.
- 115 -
NIBP Monitoring
What to do
NIBP
module,
engineer
if
required,
engineer
again,
if
failure
again,
if
failure
notify
or
notify
or