EDAN iM8 Series User Manual page 171

Patient monitor
Hide thumbs Also See for iM8 Series:
Table of Contents

Advertisement

Patient Monitor User Manual
Initialization time
Response time
Calibration
Alarm
Apnea Alarm Delay
Interfering Gas and Vapor Effect on EtCO
Gas or vapor
Nitrous oxide
Halothane
Enflurane
Isoflurane
Sevoflurane
Xenon
Helium
Desflurane
Barometric Pressure on EtCO
Quantitative effect
Ambient Barometric, Operational
0 – 40 mmHg: ± 1 mmHg additional error
41 – 70 mmHg: ± 2.5% additional error
71 – 100 mmHg: ± 4% additional error
It displays the value within 15s and meets the
requirement for measurement accuracy within 2min.
(Mainstream)
It displays the value within 20s and meets the
requirement for measurement accuracy within 2min.
(Sidestream)
60ms (Mainstream)
Not required.
EtCO
, FiCO
2
10, 15, 20, 25, 30, 35, 40s; 20s by default
Measurement Values:
2
Gas level (%)
Quantitative effect/Comments
60
Dry and Saturated Gas
4
0 – 40 mmHg: ± 1 mmHg additional error
5
41 – 70 mmHg: ± 2.5% additional error
5
71 – 100 mmHg: ± 4% additional error
5
101 – 150 mmHg: ± 5% additional error
80
*Additional worst case error when compensation
for P
50
correctly selected for the actual fractional gas
15
constituents present.
Desflurane:
The presence of desflurane in the exhaled breath at
concentrations greater than 5% will positively bias
Carbon Dioxide values by up to an additional 3
mmHg at 38mmHg.
Xenon:
The presence of Xenon in the exhaled breath will
negatively bias Carbon Dioxide values by up to an
additional 5 mmHg at 38mmHg.
Measurement Values:
2
- 162 -
and AwRR alarm
2
, O
, N
O, anesthetic agents, or helium is
B
2
2
Specifications