Monitoring Procedure - EDAN iM8 Series User Manual

Patient monitor
Hide thumbs Also See for iM8 Series:
Table of Contents

Advertisement

Patient Monitor User Manual
9
The patient monitor will be damaged if the water quantity in the water trap reaches
the limit.
10 The machine will be damaged if any pipeline from the CO
disconnected, or the air tube/air inlet/air outlet has been plugged by water or other
materials.
NOTE:
After the Low battery alarm is activated, please do not start the CO
monitor may be turned off for the low capacity of battery.

17.2 Monitoring Procedure

The principle of CO
4.3μm infrared ray. Absorption intensity is proportional to CO
the CO
concentration will compute according to the detecting CO
2
sample. The relation between partial pressure and percentage of CO
below:
LoFlo CO
module setup
2
NOTE:
You must perform a zero calibration as described in this procedure each time the ambient
temperature changes more than 10°C (for example during transport).
measurement is primarily based on the fact that CO
2
P (mmHg) = Percentage (%) × Pamp (Ambient Pressure)
Figure 17-1 LoFlo CO
WARNING
module
2
- 135 -
CO
Measuring (Optional)
2
module has been
2
measurement, or the
2
molecule can absorb
2
concentration of patient sample,
2
absorption intensity of patient
2
concentration is given
2