1
Plug the sensor cable into the monitor's CO
for warm-up.
2
Connect the cannula, airway adapter, or sample line as appropriate, to the sensor. It will
click into place when seated correctly.
3
To zero the sensor:
- Expose the sensor to room air and keep it away from all sources of CO
ventilator, the patient's breath and your own.
- Start up CO
- In the CO
SETUP menu, select ZERO CAL.
2
- The messages indicate: "zero started" , "zero successful". After the zero calibration is
finished, the user can start CO
4
For intubated patients requiring an airway adapter;
For non-intubated patients: Place the nasal cannula onto the patient.
Figure 17-2 Connecting LoFlo module
SETUP, and change WORK MODE from STANDBY to MEASURE.
2
Monitoring.
2
Figure 17-3 Air adapter
input connector. Allow the sensor two minutes
2
- 136 -
CO
Measuring (Optional)
2
including the
2