EDAN iM8 Series User Manual page 94

Patient monitor
Hide thumbs Also See for iM8 Series:
Table of Contents

Advertisement

Patient Monitor User Manual
Users can access the following functions:
SMART LEAD OFF: in 5 LEADS mode, if the CH1 and CH2 can not measure
because of the lead off or other reasons, it can shift to other LEADS to collect a ECG
waveform.
BEAT VOL
Six selections are available: 0, 1, 2, 3, 4, 5. "5" indicates maximum volume. "0"
indicates no sound.
PACE
detected signal will be marked by a "
ON
for non-paced patient.
OFF
NOTE:
When monitoring a patient with a pacemaker, set "PACE" to ON. If monitoring a patient
without a pacemaker, set "PACE" to OFF.
If "PACE" is ON, the system will not perform some types of ARR analysis. For detailed
information, please refer to the Section ARR ALARM.
CASCADE: turn on or off CASCADE display. When it is set to ON, the ECG
waveform is display in 2 channels. This function is available only for the NORMAL
DISPLAY in ECG DISPLAY.
ECG CAL
Pick this item to start ECG calibrating process. Picking this item again can finish calibrating
process.
Users can turn on or off the Power frequency filter of 50Hz or 60Hz in DIAGNOSTIC
mode. It can use standardized voltage to set the display width for 1mV signal. For example
X1 is for 10mm, X2 is for 20mm.
ADJUST WAVE POS
Figure 12-8 Other Setup menu
- 85 -
ECG/ RESP Monitoring
|
" above the ECG waveform