Used to adjust the position of ECG waveform on the screen, pick this item to call up the
ADJUST WAVE POS dialog box. The user may use CH NAME item to select the channel
to be adjusted, UP-DOWN to adjust the position of the selected channel on the screen,
BACK TO DEFAULT to let the waveform go back to the default position on the screen.
DEFAULT
Pick the DEFAULT item to call up the ECG DEFAULT CONFIG dialog box, in which
you can select the FACTORY DEFAULT CONFIG or the USER DEFAULT CONFIG
item. After selecting any of the items and exiting the dialog box, the system will pop up a
dialog box asking for your confirmation.
For patients with pacemakers, the pacing impulse analysis function must be switched ON.
Otherwise, the pacing impulse may be counted as normal QRS complex, which results in
failure of ECG LOST error detection.
NOTE:
When PACE Switch is ON, the Arrhythmia events related to PVCs will not be monitored.
At the same time, the ST analysis will not be performed either.
If the monitor can do ST segment monitoring and Arrhythmia monitoring, please refer to Section
12.7 and 12.8.

12.6 ECG Alarm Information

Alarms occurring in the process of ECG measurement contain two types: physiological alarm and
technical alarm. For the audio and visual features during the appearance of these alarms in the
process of ECG measurement, please refer to the related description in Chapter6 Alarm. On the
screen, physiological alarm messages are displayed in the Physiological Alarm area. Technical
alarms messages are displayed in the Technical Alarm area. This section does not describe the
content about Arr. and ST analysis.
Tables below describe respectively the possible various alarms that may occur during the
measurement.
Figure 12-9 ADJUST WAVE POS Menu
WARNING
- 86 -
ECG/ RESP Monitoring