Maintenance And Cleaning - EDAN iM8 Series User Manual

Patient monitor
Hide thumbs Also See for iM8 Series:
Table of Contents

Advertisement

Patient Monitor User Manual
Message
There is interference with
SpO
Noisy
SpO
2
2
Signal
and
abnormal.
Prompt message:
SpO
SpO
2
SEARCH
disconnected
PULSE
patient or the monitor.

13.6 Maintenance and Cleaning

1
Before cleaning the monitor or the sensor, make sure that the equipment is switched
off and disconnected from the power line.
2
Do not subject the sensor to autoclaving.Do not immerse the sensor into any
liquid.Do not use any sensor or cable that may be damaged or deteriorated.
Use a cotton ball or a soft mull moistened with hospital-grade ethanol to wipe the surface of
the sensor, and then dry it with a cloth. This cleaning method can also be applied to the
luminotron and receiving unit.
The cable can be cleaned with 3% hydrogen dioxide, 70% isopropanol, or other active
reagent. However, connector of the sensor shall not be subjected to such solutions.
Cause
measurement signals
the
waveform
sensor
may
2
from
Alarm Level
Low
is
be
Low
the
WARNING
- 107 -
SpO
2
What to do
Check the condition of patient
and avoid patient movement;
make sure the cable is well
connected.
If the Pulse is not displayed
after
30s,
check
connection
between
sensor
and
the
finger, reconnect the sensor or
change it to other parts.
Monitoring
the
the
patient's

Advertisement

Table of Contents
loading

Table of Contents