A1.2.6 Recall

Trend Recall
Recall

A1.2.7 ECG

Lead Mode
Waveform
Lead naming style
Display Sensitivity
Sweep
Bandwidth (-3dB)
CMRR
(Common
Rejection Ratio)
Notch
Differential Input Impedance
Input Signal Range
Accuracy
of
Input
Reproduction
Electrode
Offset
Tolerance
Auxiliary Current (Leads off
detection)
1 hrs, 1-second resolution
96 hrs, 1-min. resolution
500 sets NIBP measurement data
3-Lead: I, II, III
5-Lead: I, II, III, aVR, aVL, aVF, V
3-Lead: 1-channel waveform
5-Lead: 2-channel waveform, max. seven waveforms;
AHA, IEC
1.25mm/mV (×0.125), 2.5mm/mV (×0.25), 5mm/mV
(×0.5), 10mm/mV (×1), 20mm/mV (×2), 40mm/mV (×4),
AUTO gain
6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
Diagnosis: 0.05Hz to 150Hz
Monitor: 0.5Hz to 40Hz
Surgery: 1Hz to 20Hz
Diagnosis: >95dB (the Notch filter is off)
Mode
Monitor: >105dB (the Notch filter is on)
Surgery: >105dB (the Notch filter is on)
50Hz/60Hz (Notch filter can be turned on or off manually)
>5MΩ
±10mV PP
Signal
The total error and frequency response comply with
ANSI/AAMI EC13:2002, Sect. 4.2.9.8.
Potential
±500mV
Active electrode: <100nA
Reference electrode: <900nA
- 153 -
Specifications