EDAN iM8 Series User Manual page 163

Patient monitor
Hide thumbs Also See for iM8 Series:
Table of Contents

Advertisement

Patient Monitor User Manual
Recovery
time
Defibrillation
Leakage current of patient
Scale signal
System noise
ESU Protection
Noise
Suppression
Electrotome
Pace Pulse
Pulse indicator
Pulse Rejection
Minimum input slew rate
Heart rate
Range
Accuracy
Resolution
Sensibility
PVC
Range
Resolution
ST value
after
<5s
<10μA
1mVPP, accuracy is ±5%
<30μVPP (RTI)
Restore time: ≤10s
Meets the requirements of ANSI/AAMI EC13-2002: Sect.
4.1.2.1 a)
of
Tested according to the test method in EC13: 2002
Sect.5.2.9.14, it accords with the standard.
Pulse is marked if the requirements of ANSI/AAMI
EC13:2002, Sect. 4.1.4.1 are met:
Amplitude: ±2 mV ~ ±700 mV
Width: 0.1 ms ~2.0 ms
Ascending time: 10 μs ~ 100 μs
Pulse is rejected if the requirements of ANSI/AAMI
EC13-2002: Sect. 4.1.4.1 are met:
Amplitude: ±2 mV ~ ±700 mV
Width: 0.1 ms ~2 ms
Ascending time: 10 μs ~100 μs
>2.5V/S
ADU: 15 bpm ~ 300 bpm
PED/NEO: 15 bpm ~ 350 bpm
±1% or 1 bpm, whichever is greater
1 bpm
≥300 μVPP
ADU: 0~300 PVCs/ min
PED/NEO: 0~350 PVCs/ min
1 PVCs/min
- 154 -
Specifications