Patient Monitor User Manual
REC WAVE: it can be set to any waveform of 8s, such as IBP1, CO
also be set to OFF.
RECALL: Used to review frozen waveforms.
REC: select this item to record the setting waveform in REC WAVE.
EXIT: After this button is pressed, the system closes the FROZEN menu and exits the
Freeze status.
NOTE:
Pressing the FREEZE button repeatedly in a short period of time may result in
discontinuous waveforms displaying on the screen.

7.4 Reviewing Frozen Waveform

By moving the waveform, you may review a waveform of 12 minutes before it is frozen. For a
waveform of less than 12 minutes, the remaining part is displayed as a straight line. Use the
rotary snob on the control panel to move the cursor to the RECALL option in the FROZEN
menu. Press the knob. By turning the knob left or right, frozen waveforms on the screen will
move left or right correspondingly. There is an arrow indicating upward on the right side of the
last waveform. There is also a time scale beside the arrow. "-0S" is used to mark the moment
when waveforms are frozen. With waveforms moving right, this time mark will in turn change
into -1S, -2S, -3S... These time marks are applied to all waveforms on the screen.
Figure 7-1 Frozen
- 55 -
Freeze