Attaching the objective
40
1
Screw the objective into the mount hole of the revolving nosepiece
a
When changing the objective, hold the revolving nosepiece a and
rotate it so that the intended objective comes exactly above the
specimen.
NOTE
TIP
Removing the objective
1
If the objective is screwed in too firmly, use the slip prevention rubber
sheet, etc. to rotate the objective in the direction opposite to the
attaching direction (arrow direction).
a carefully not to drop it.
· Do not rotate the revolving nosepiece by holding the
objective.
· Be careful if you rotate the revolving nosepiece while
observing the edge of the slide glass with the high
magnification objective (40X, etc.), the objective may
interfere with the specimen holder.
· Only objectives of 24mm in diameter can be attached to all 5
holes of the revolving nosepiece.
· For available objectives by observation method, refer to the
table "Available observation methods by objective" on page 35.