Setting The Measured Value; Correcting Value - ProMinent DULCOMETER D1Cb Assembly And Operating Instructions Manual

Hide thumbs Also See for DULCOMETER D1Cb:
Table of Contents

Advertisement

Measured Variables and Operating Menus for Amperometric Sensors
7.7

Setting the Measured value

WARNING!
Incorrect metering due to incorrect metering range
Possible consequence: Fatal or serious injuries.
The measuring range of the sensor is essential for the measuring range!
If the assignment of the measuring range is modified, the settings must be
checked in all menus
If the assignment of the measuring range is changed, the sensor must be recali‐
brated
The relevant information can be found in the operating manual for the sensor/
measuring equipment
measured value
setting ?
Fig. 39: Measured value
This setting is solely for matching the DULCOMETER
third party suppliers. Sensors provided by third party providers may have measuring
ranges, which deviate from the default settings of the DULCOMETER
To match a sensor from Prominent to the DULCOMETER
use the menu under
Variable and Measuring Range' on page 54
7.8

Correcting value

Only necessary when using the DULCOTEST
ClO
.
2
Standard signal input mA
range adjustment
measured value
4 m A
=
0.00 ppm
'General Settings' , see Ä Chapter 5.1.2 'Selection of the Measured
range adjustment
measured value
20 mA =
2 .00 ppm
®
D1Cb / D1Cc to the sensors of
D1Cb / D1Cc, please only
®
CDP sensor for chlorine dioxide
®
A0019_GB
D1Cb / D1Cc.
®
79