9 Calibration
L
C
OAD
ALIBRATION FROM
Unit [n] > Calibration > Load from File
Use to load a previously saved calibration file.
It is recommended that the current calibration settings be saved using Unit [n] > Calibration >
Save to File prior to loading another calibration file. When loading a previously saved calibration
file, a backup of the current file is created and saved as [SN]last.cal. The backup file is overwritten
each time a new one is created.
Changing the calibration may affect the analyzer's performance.
S
C
AVE
ALIBRATION TO
Unit [n] > Calibration > Save to File
Use to save the current calibration settings to a backup file which can later be reloaded using the
Unit [n] > Calibration > Load from File menu option.
The default file naming convention for calibration files can be used or the file name can be
changed. The default file name of 0217-2013-04-25.CAL is interpreted as:
0217
Analyzer serial number
2013-04-25
Date the calibration file was saved
.CAL
File name extension