Unit [n] > Calibration > Sample Servo
Unit [n] > Calibrate Analysis > Sample Servo
Use to calibrate the servo valve to the sample transducer. The servo valve should always be
recalibrated after a pressure calibration has been performed. The pressure transducer should be
calibrated before starting this calibration procedure.
Ensure the pressure transducer has been calibrated before performing this pro-
cedure. Go to Unit [n] > Unit Configuration and view the calibration information.
Contact your Micromeritics Service Representative if calibration dates are not listed.
Click Start. The window closes when the calibration is complete.
Click
Cancel
to stop the calibration process.
Gemini VII Operator Manual
239-42828-01 (Rev H ) — May 2021
V
ALVE
Sample Servo Valve
9 - 5