Emergency Button - Planmeca Viso Technical Manual

Hide thumbs Also See for Viso:
Table of Contents

Advertisement

1.4.2.1 Calibrating touch screen display

1.4.3 Emergency button

Technical manual
If you want to pause a function (instead of cancelling it) touch the pause
button.
To scroll a list down or up, slide your finger on the screen.
To change a setting, select the service spanner at the top left corner of the
main view. This will take you to the settings menu where you can adjust the
settings of the X-ray unit.
In case the touch screen is not working accurately, e.g. touching button on
the screen activates task of the button next to it, or it is just hard to activate
desired button or area on the screen, calibrate the display.
Normally, there is no need to calibrate the display at all, but the factory made
calibration remains valid.
To calibrate the display, follow the steps below.
1. Switch on the X-ray unit.
2. As soon as the screen displays text Touch the screen now to start
calibration, touch the screen. The calibration screen appears.
3. In the calibration screen, five calibration adjustment markers will appear
on the display, one at a time:
1st in the left up corner
2nd in the right up corner
3rd in the right bottom corner
4th in the left bottom corner
5th in the center of the display.
4. Touch the adjustment markers in the order they appear on the screen.
After touching the final adjustment marker, the X-ray unit will start up
normally.
The emergency stop button is located on the top of the stationary column.