5 Adjustments and calibrations
82
Planmeca Viso
2. Click Next on the Geometry - Elbow arm symmetrical window. The
system will move to Planmeca QA - QA PM C-arm center calibration.
3. Protect yourself from radiation and press the exposure button to take the
exposure. The test results will appear in the window.
4. If the QA test fails, the unit is not correctly adjusted and calibrated.
Check the patient support lift calibration. Refer to section "Patient
support lift calibration" on page 103.