Pose volume calibration
Technical manual
1. Attach the surface phantom to the adapter on the patient support base.
2. Click Next on the Camera calibration - chessboard calibration window.
The system will move to Camera calibration - Pose volume calibration.
3. Protect yourself from radiation and press the exposure button to take the
exposure. The esposure results will appear in the window.
5 Adjustments and calibrations
Planmeca Viso
85