Planmeca Viso Technical Manual page 45

Hide thumbs Also See for Viso:
Table of Contents

Advertisement

E240
E241
E242
E243
E244
Error
Explanation
code
E245
Overcurrent
E246
E251
Directional error
E252
E256
E267
Patient support error Patient support moved during the
Technical manual
Overcurrent detected in collimator Y
motor which is connected to CAM
PCB J25 and controlled by MCM
module 5.
Overcurrent detected in tube head
rotation motor which is connected to
CAM PCB J22 and controlled by
MCM module 6.
Overcurrent detected in collimator
filter revolver motor which is
connected to CAM PCB J22 and
controlled by MCM module 6.
Overcurrent detected in collimator
filter motor which is connected to
CAM PCB J20 and controlled by
MCM module 7.
Overcurrent detected in collimator X2
left motor which is connected to CAM
PCB J20 and controlled by MCM
module 7.
Overcurrent detected in collimator Y2
lower motor which is connected to
CAM PCB J25 and controlled by
MCM module 7.
Overcurrent detected in tube head
rotation motor which is connected to
CAM PCB J31 and controlled by
MCM module 8.
Directional error at shoulder motor
which is connected to PSU PCB J7
and controlled by MCM module 1.
Directional error at elbow motor which
is connected to PSU PCB J7 and
controlled by MCM module 2.
Directional error at c-arm motor which
is connected to CAM PCB J21 and
controlled by MCM module 4.
patient support locking mechanism.
3 Error messages
Comments
The motor drives the
mechanism into the wrong
position. Check the sensors
and the cabling.
The motor drives the
mechanism into the wrong
position. Check the sensors
and the cabling.
The motor drives the
mechanism into the wrong
direction. Check the
sensors and the cabling.
Planmeca Viso
39