E291
Collimator
E292
E293
E294
E295
E296
E297
E298
Tube head rotation
E299
Z lock
Error
Explanation
code
E301
Filament voltage
E302
E303
Error
Explanation
code
E311
Tube voltage
E312
E313
E317
E318
Tube voltage
E319
Tube voltage
Technical manual
Collimator right X self test failed.
Collimator left X self test failed.
Collimator upper Y self test failed.
Collimator lower Y self test failed.
Collimator filter revolver self test
failed.
Collimator X collision detected.
Collimator Y collision detected.
Tube head rotation self test failed
Z lock test failed
X-ray generation related errors (E3xx)
The following table lists the X-ray generation related errors.
Filament voltage missing
completely.
Filament voltage too low during
preheat.
Filament voltage too high during
preheat.
Tube voltage missing completely.
Tube voltage too low.
Tube voltage too high.
Tube voltage overshot suddenly
without arching.
Tube mA overshot suddenly
without arching.
Tube pulse length exceeded the
requested value.
3 Error messages
Comments
Exposure not possible. Check
the cabling, interfaces and
PSU- PCB and tube head.
Exposure not possible. Check
the cabling, interfaces and
PSU- PCB and tube head.
Exposure not possible. Check
the tube head and the PSU-
PCB.
Comments
Exposure interrupted.
Exposure not possible. Check
the tube head and the PSU
PCB.
Exposure not possible. Check
the prefilament values are
correct or perform the
prefilament calibration again.
Exposure interrupted. Check
the interface, cabling and the
PSU PCB.
Exposure interrupted. Check
the functionality of the tube
head and the PSU PCB.
Planmeca Viso
41