Planmeca Viso Technical Manual page 105

Hide thumbs Also See for Viso:
Table of Contents

Advertisement

Technical manual
3. Drive the patient support with the height adjusting buttons so that the
carriage block lower edge (white line in the figure below) is in line with
sensor center point.
4. Check the Patient support lift field:
If the value in the field is green, the sensor is correctly calibrated.
If the value in the field is red, the sensor has to be calibrated, see the
next step.
5. If needed, calibrate the patient support position sensor.
6. Loosen the two screws at the side of the cog wheel of the sensor with 2
mm Allen key.
7. Rotate the sensor until the value in the Patient support sensor field turns
green (close to zero) in the Patient Support display.
5 Adjustments and calibrations
Planmeca Viso
99