5.9 Patient support lift calibration
Technical manual
1. Tighten the adjustment screw by hand (1) until the spring inside is fully
depressed.
The spring depresses to its full extent and the adjustment screw cannot
be easily tightened by hand.
CAUTION
Do not tighten using an Allen key or other tools, as over-tightening will cause
damage to the friction pad or the screw threads.
1. Release the adjustment screw by a few turns, until the screw turns more
freely but there is still some tension on the screw.
2. Tighten the locking nut by hand (2) against the mechanism base to hold
the adjustment screw's position.
NOTE
Before calibrating the patient position mechanism, the patient support sensor
calibration must be done, see instructions in section
1. Select Technical from the bottom of the display.
The technical setting mode is password protected and the password is
asked when the mode is entered for the first time after switching the unit
on. The password is 1701.
5 Adjustments and calibrations
Planmeca Viso
103