6 Parts replacement
Removing X-ray tube head rear cover
Removing C-arm bottom cover
128
Planmeca Viso
1. Rotate the X-ray tube so it is flush with the C-arm (1).
2. Loosen the four attachment screws on the bottom and sides of the rear
cover (2). Remove the cover from the X-ray tube head (3 and 4).
Remove the following covers before removing the C-arm botton cover:
1. C-arm top plate
2. C-arm top cover
3. Sensor rear cover
4. Tube head front cover
1. Bend the rim outwards slightly to aid in removal (1).
2. Unscrew the four screws that attach the bottom cover to the C-arm (2).
3. Rotate the X-ray tube head so it is flush with the C-arm.