Emergency Stop Button - Planmeca ProMax User Manual

Cephalostat
Hide thumbs Also See for ProMax:
Table of Contents

Advertisement

3 Main parts

3.4 Emergency stop button

6
Planmeca ProMax
The emergency stop button is located on the top of the stationary column.
Press the button to stop the X-ray unit operating in an emergency. When the
emergency stop button is pressed down, all movements of the X-ray unit are
blocked and the unit will not generate radiation. The up / down movement
will stop within a distance of 10 mm (0.4 in.).
A help message will appear on the touch screen. Guide the patient away
from the X-ray unit. Then release the emergency stop button. The X-ray unit
will automatically restart.
User's manual

Advertisement

Table of Contents
loading

Table of Contents