Planmeca Viso Technical Manual page 23

Hide thumbs Also See for Viso:
Table of Contents

Advertisement

Patient supports
Patient handles
Generator (according to IEC 60601‐2‐7: 1998)
X‐ray tube
Focal spot size (according to IEC 60336: 2005)
Filtration
3D
Pan (SmartPan) / ProCeph
Tube housing front cover quality equivalent
filtration (not included in the specified total
filtration)
Anode voltage
3D
Pan (SmartPan)
ProCeph
Anode current
3D
Pan (SmartPan)
ProCeph
mAs range
Dose range and accuracy
Linearity of radiation output
Exposure time
3D
Pan (SmartPan)
ProCeph
SID
3D / Pan (SmartPan)
Ceph
Technical manual
1 General and technical data
As shown in section Patient supports in User
Manuals
Resonant-mode, DSP-controlled, 80 ‐ 160 kHz
Toshiba D-059SBR or SXR 130-10‐0.5 SC
0.5 x 0.5 mm
Total 2.5 mm Al + 0.2 mm / 0.5 mm Cu
Total 2.5 mm Al
0.3 mm Al @ 70 kV / HVL 2.6 mm Al
80 ‐ 120 kV ±5%
60 ‐ 84 kV ±5%
60 ‐ 84 kV ±5%
Toshiba D-059SBR: 1‐12.5 mA ±10%
SXR 130‐10‐0.5 SC: 1‐16 mA ±10%
Toshiba D-059SBR: 1‐14 mA ±10%
SXR 130‐10‐0.5 SC: 1‐16 mA ±10%
Toshiba D-059SBR: 14 mA ±10%
SXR 130-10‐0.5 SC: 16 mA ±10%
min. / max. as indicated ±(10% + 0.2 mAs)
Dose range min. / max. as indicated on system
user interface.
Accuracy of dosimetric indication (DAP, CTDI):
±40%
< 0.1
Pulsed, effective 1.5 ‐ 36 s as indicated ±10%
2.5 – 15.6 s as indicated ±10%
0.1 – 1.6 s as indicated ±10%
700 mm
1700 mm (66.9 in.)
Planmeca Viso
17