5.3.3.2 Flat field 3D calibration
Technical manual
2. Protect yourself from radiation and press the exposure button to take the
exposure. The exposure results will appear in the window.
3. Perform the Collimator adjustment - FOV BOTTOM, LEFT and RIGHT
calibrations.
1. Click Next on the Collimator adjustment - FOV RIGHT window. The
system will move to Flat field 3D 1x1 3pFcalibration.
5 Adjustments and calibrations
Planmeca Viso
71