Planmeca Viso Technical Manual page 14

Hide thumbs Also See for Viso:
Table of Contents

Advertisement

1 General and technical data
8
Planmeca Viso
Press the emergency button to stop the X-ray unit operating in an
emergency. When the emergency stop button is pressed down, all
movements of the X-ray unit are blocked and the unit will not generate
radiation. The up / down movement will stop within a distance of 10 mm (0.4
in.).
A help message will appear on the touch screen. Guide the patient away
from the X-ray unit. Then release the emergency stop button. The X-ray unit
will automatically restart.