3 Error messages
E321
Tube current
E322
E323
E332
Arcing
E334
E336
E342
Monitoring errors
E343

3.2 Feedback errors (E4xx)

Error
Explanation
code
42
Planmeca Viso
Tube current missing completely.
Tube current too low.
Tube current too high.
Severe arcing across x-ray tube.
Severe arcing at tube head anode
end.
Severe arcing at tube head
cathode end.
Exposure pulsing fault.
Conflict between assembled and
configured X-ray tube type.
The following table lists the feedback errors.
Exposure interrupted. Check
the interface, cabling and the
PSU PCB.
Exposure interrupted. Check
the interface, cabling and the
PSU PCB.
Exposure interrupted. Perform
the preheat calibration again.
Check the interface, cabling
and the PSU PCB.
Exposure interrupted.
Check the tube head
functionality. If the X-ray unit
has not been used for a week
or more, take four test
exposures with a one-minute
interval between exposures.
Use the lowest possible kV
and mA values and a long
exposure time.
Exposure interrupted.
Exposure interrupted.
Exposure not possible.
Check that correct type of tube
head is selected in the 3100
Unit Configuration menu, in
Tube Type setting.