6 Parts replacement
Removing sensor back cover
126
Planmeca Viso
Remove the C-arm top plate before removing the sensor back cover.
1. The sensor back cover is attached to the sensor frame with six
attachment screws (1).
2. Loosen the two bottom screws approximately three turns, but do not
unscrew completely (2).
3. Loosen the two screws on the inside (front) surface of the sensor (3).
4. Loosen the two top screws approximately three turns, but do not
unscrew completely (4).
5. Remove the back cover (5).