5 Adjustments and calibrations
5.8 Patient support arm adjustments
5.8.1 Calibrating patient support position sensor
98
Planmeca Viso
3. Set the desired maximum driving speed of column motor movement with
the arrow buttons.
4. Exit the calibration mode by touching the green check mark button.
The patient support position sensor must be calibrated after the sensor is
replaced.
1. Enter the calibration mode:
1. Touch the service spanner on the Main display.
2. Touch Technical and enter password (1701).
3. Touch Calibrations (3200) and select Patient support lift (3240) and
Patient support lift position sensor (3241).
2. The following display appears.
Technical manual