Original Manufacturer; User's Statement - Planmeca Viso Technical Manual

Hide thumbs Also See for Viso:
Table of Contents

Advertisement

Air pressure
Operating:
Temperature
Relative humidity
Air pressure
Max. altitude
Image properties
ProCeph:
Flat panel pixel size
Flat panel active surface
3D:
Flat panel pixel size
Flat panel active surface
Pan (SmartPan):
Flat panel pixel size
Flat panel active surface
Operating requirements for ProFace program
Optimum colour temperature
Even and uniform lighting
No bright lights

1.6.2 Original manufacturer

1.7 User's statement

User's statement information for Planmeca Viso
Feature / property
Radiation leakage technique factors
Filtration
Maximum attenuation equivalent of the front
panel of the 3D sensor head
Technical manual
Planmeca Oy, Asentajankatu 6, FIN-00880, Helsinki, Finland
Phone: +358 20 7795 500, Fax: +358 20 7795 555,
This section contains the user's statement information for Planmeca Viso
equipped with 120 kV tube head (Toshiba D-059SBR or Superior SXR
130-10-0.5 SC). Values differing between the tube head types are
separately presented for each tube head type.
1 General and technical data
700 ‐ 1060 hPa
+10°C ‐ +35°C (+50°F ‐ +95°F)
10 ‐ 90% RH (non-condensing)
800 ‐ 1060 hPa
2000 m (1.25 miles)
139 µm
302 x 249 mm (11.89 x 9.80 in.)
139 µm
299.7 x 246.3 mm (11.80 x 9.70 in.)
139 µm
8 ‐ 25 x 146 mm (0.31 ‐ 0.98 x 5.74 in.)
Approx. 6500 Kelvin
Details
Toshiba D-059SBR or Superior SXR 130-10-0.5
SC: The maximum rated peak tube potential is
120 kV and the maximum rated continuous tube
current is 2.1 mA for the maximum rated peak
tube voltage.
The Radiation port contains additional filtration of
0.2 mm or 0.5 mm Cu. When the X-ray beam is
attenuated with the 4.3 mm Al (at 120kV) the
resulting dose is 0.5 - 0.7 times the original.
1.2 mmAl
www.planmeca.com
Planmeca Viso
19