6 Panoramic and 3D calibrations and adjustments
66
6.5 The calibration sequence
6.5.1
Calibration of the preheat of the tube
NOTICE! Calibration of the preheat of the tube takes a
while. The audible signal starts and stops several times.
Do this calibration if there has been long time between
installation and final testing of the unit or if the other
calibrations won't pass.
1. Select the Preheat program.
2. Take an exposure.
3. Touch screen display and s2terminal inform you when
the calibration is done. Repeat the calibration until cali-
bration result "passed" is achieved.
6.5.2
Calibration of the tube current
NOTICE!
Calibration
approximately 10 to 15 minutes. The audible signal starts
and stops several times during the calibration.
Do this calibration if there has been long time between
installation and final testing of the unit.
1. Select the mA program.
2. Take an exposure.
3. Touch screen display informs when the calibration is
done. Repeat the calibration until calibration result
"passed" is achieved.
of
the
tube
current
takes