Recommended Safe Distances Between Portable And Mobile Hf Telecommunications Equipment And The Mastersurg Lux Wireless - KaVo MASTERsurg LUX Wireless Instructions For Use Manual

Table of Contents

Advertisement

Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
12 Information about electromagnetic compatibility | 12.4 Recommended safe distances between portable and mobile HF
telecommunications equipment and the MASTERsurg LUX Wireless
Interference immunity
tests
Wire-based HF interfer-
ence according to IEC
61000-4-6
Wireless HF interfer-
ence according to IEC
61000-4-3
IEC 60601 test levels
3 V
eff
150 kHz to 80 MHz
3 V/m
800 MHz to 2.5 GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not be applicable in every case. The propagation of
electromagnetic waves is subject to absorption and reflection by buildings, objects,
and people.
a
The field strength of stationary transmitters such as base stations of mobile tele-
phones and land mobile radio devices, amateur radio stations, AM and FM radio and
television broadcasting stations cannot be determined based on theoretical considera-
tions. A site study should be considered to determine the electromagnetic environment
in terms of stationary transmitters. If the field strength measured at the location at
which the MASTERsurg LUX Wireless is used exceeds the conformity levels specified
above, the MASTERsurg LUX Wireless should be monitored to confirm that it is func-
tioning as intended. Should unusual performance features be observed, additional
measures may be required, such as, e.g., a different alignment or another location for
the MASTERsurg LUX Wireless.
b
In the frequency range of 150 kHz to 80 MHz, the field strength should be less than
3V
V/m.
eff
12.4 Recommended safe distances between portable and mobile HF
telecommunications equipment and the MASTERsurg LUX Wireless
The MASTERsurg LUX Wireless is intended for use in an electromagnetic environ-
ment like the one specified below. The customer or user of the MASTERsurg LUX
Wireless can help to avoid electromagnetic interference by complying with the min-
Compliance level
3 V
eff
3 V/m
72 / 76
Electromagnetic environment -
Guidelines
Portable and mobile radio devices
should not be used closer to the
MASTERsurg LUX Wireless includ-
ing the cables than the recommen-
ded safe distance calculated using
the equation for the transmission
frequency.
Recommended safe distance:
d = [3.5/3]
P
= 1.17
P
d= [3.5/3]
P
=1.17
P
for 80 MHz to
800 MHz
d= [7.0/3]
P
= 2.33
P
for 800 MHz
to 2.5 GHz
where P is the maximal nominal
power of the transmitter in watts
(W) as specified by the transmitter
manufacturer and d is the recom-
mended safe clearance in metres
(m).
The field strength of stationary wire-
less radio transmitters as measured
a
locally
should be lower than the
conformance level at all frequen-
b
cies.
Interference is possible in the vicin-
ity of devices that bear the following
symbol.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1.009.1200

Table of Contents