10 Information concerning the electromagnetic compatibility according to EN IEC 60601-1-2 | 10.4 Recommended
safe distances between portable and mobile HF telecommunications equipment and this product
Rated power P of the
transmitter in W
0.01
0.1
1
10
100
U1 = Compliance level according to 4-6: 3 V
E1 = Compliance level according to 4-3: 3 V/m
Factor
d
Within the frequency range of 150 kHz to 80 MHz, the field strength should be
less than 1 V/m.
10.4 Recommended safe distances between portable
and mobile HF telecommunications equipment and this
product
This product is intended for use in an electromagnetic environment in which HF
disturbances are controlled. The customer or user of the product can help pre-
vent electromagnetic interference by maintaining minimum distances between
portable and mobile high-frequency communication devices (transmitters) and
the product as recommended below corresponding to the maximum output of
the communication device.
Safe distance depending on the transmission frequency in m
150 kHz to 80 MHz
d=1.17
P
0.12
0.37
1.17
3.69
11.7
eff
[3.5/U1]
For transmitters whose maximum rated power is not included in the above
table, the recommended safe distance d in metres (m) can be calculated using
the equation for the respective column, where P is the maximum rated power
of the transmitter in Watts (W) as specified by the manufacturer.
NOTE: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not be applicable in every case. The spread of
electromagnetic waves is absorbed and reflected by buildings, objects and
people.
80 MHz to 800 MHz
d=1.20
P
0.12
0.38
1.20
3.79
12
[12/E
]
1
176 / 178
800 MHz to 2.5 GHz
d=2.3
P
0.23
0.73
2.3
7.27
23
[23/E
]
1