PROline Prosonic Flow 93 FOUNDATION Fieldbus
Medium temperature
range
Limiting medium pressure
range
(nominal pressure)
Pressure loss
Design, dimensions
Weight
Material
Endress+Hauser
10.1.9
Operating conditions (process)
• Flowrate measuring sensors Prosonic Flow P (Clamp On):
–40...+80 °C / 0...+170 °C
• Flowrate measuring sensors Prosonic Flow W (Clamp On):
–20...+80 °C
• Flowrate measuring sensors Prosonic Flow W (Insertion):
–40...+80 °C
• Sound velocity measuring sensors DDU 18:
–40...+80 °C / 0...+170 °C
• Wall thickness measuring sensor DDU 19:
0...+60 °C
Perfect measurement requires that the static fluid pressure is higher than vapour
pressure.
Maximum nominal pressure for W sensors (Insertion): PN 16 (PSI 232)
There is no pressure loss.
10.1.10 Mechanical construction
see Page 116 ff.
Transmitter wall-mount housing: 6.0 kg
Measuring sensors:
• Flowrate measuring sensors P incl. mounting rail and tensioning band: 2.8 kg
• Flowrate measuring sensors W (Clamp On) incl. mounting rail and tensioning bands:
2.8 kg
• Flowrate measuring sensors W (Insertion / single path version): 4.5 kg
• Flowrate measuring sensors W (Insertion / dual path version): 12.0 kg
• Sound velocity measuring sensors DDU 18 incl. tensioning band: 2.4 kg
• Wall thickness measuring sensor DDU 19 incl. tensioning band: 1.5 kg
Transmitter housing 93 (wall-mount housing):
Powder-coated die-cast aluminum
Standard designations of the materials (measuring sensors P / W / DDU 18 / DDU 19):
Sensor housing W (Clamp On)
Nickled brass
Sensor housing P / DDU 18 / DDU 19
Sensor holder
Welding parts for W sensors
(Insertion version)
Sensor contact surface
Tensioning bands
High temperature sensor cable
– Cable connector (stainless steel)
– Cable sheath
Standard sensor cable
– Cable connector (nickled brass)
– Cable sheath
DIN 17660
2.0401
DIN 17440
1.4301
1.4301
1.4301
Chemical-resistant plastic
1.4301
1.4301
PTFE
2.0401
PVC
10 Technical data
UNS
C38500
AISI
304
304
304
304
304
PTFE
C38500
PVC