Operating Conditions: Process - Endress+Hauser Proline Promass 84 MODBUS RS485 Operating Instructions Manual

Coriolis mass flow measuring system for custody transfer
Table of Contents

Advertisement

Proline Promass 84 MODBUS RS485
Medium temperature range
Limiting medium pressure
range (rated pressure)
Limiting flow
Endress+Hauser
11.1.9

Operating conditions: Process

Sensor
• Promass A: –50 to +200 °C (–58 to +392 °F)
• Promass F: –50 to +200 °C (–58 to +392 °F)
• Promass F (high temperature version): –50 to +350 °C (–58 to +662 °F)
• Promass M: –50 to +150 °C (–58 to +302 °F)
Seals
• Promass F:
No internal seals
• Promass M:
– Viton: –15 to 200 °C (–5 to +392 °F)
– EPDM: –40 to +160 °C (–40 to +320 °F)
– Silicone: –60 to +200 °C (–76 to +392 °F)
– Kalrez: –20 to +275 °C (–4 to +527 °F)
– FEP: sheathed (not for gas applications): –60 to +200 °C (–76 to +392 °F)
• Promass A (only for mounting sets with threaded connections):
– Viton: –15 to 200 °C (–5 to +392 °F)
– EPDM: –40 to +160 °C (–40 to +320 °F)
– Silicone: –60 to +200 °C (–76 to +392 °F)
– Kalrez: –20 to +275 °C (–4 to +527 °F)
The material load diagrams (pressure-temperature diagrams) for the process connections are
provided in the separate "Technical Information" document on the measuring instrument in
question. This can be downloaded as a PDF file from www.endress.com.
List of the "Technical Information" documents available → ä 118.
Pressure ranges of secondary containment:
• Promass F
– DN 8 to 50 (3/8" to 2"): 40 bar (600 psi)
– DN 80 (3"): 25 bar (375 psi)
– DN 100 to 150 (4" to 6"): 16 bar (250 psi)
– DN 250 (10"): 10 bar (150 psi)
• Promass M
– 100 bar (1450 psi)
• Promass A
– 25 bar (362 psi)
See the "Measuring range" section → ä 93
Select nominal diameter by optimizing between required flow range and permissible pressure loss.
See the "Measuring range" section for a list of max. possible full scale values.
• The minimum recommended full scale value is approx. 1/20 of the max. full scale value.
• In most applications, 20 to 50% of the maximum full scale value can be considered ideal.
• Select a lower full scale value for abrasive substances such as liquids with entrained solids
(flow velocity <1 m/s (<3 ft/s)).
• For gas measurement the following rules apply:
– Flow velocity in the measuring tubes should not be more than half the sonic velocity
(0.5 Mach).
– The maximum mass flow depends on the density of the gas:
Formula → ä 95
Technical data
107