Technical data
40
10.3
Power supply
Supply voltage
Power consumption
Current consumption
Fuse
10.4
Performance characteristics
Measuring interval
Measuring time
Measuring deviation
Sample requirement
Reagent requirement
Calibration interval
Flushing interval
Maintenance interval
Servicing requirements
10.5
Operating conditions (environment)
Ambient temperature
Protection class
230 V AC, 50/60 Hz
115 V AC, 50/60 Hz
non-cooled:
40 VA
cooled:
200 VA
non-cooled:
0.15 A
cooled:
0.9 A
medium time-lag 0.2 A
time-lag 0.5 A
2 ... 120 min
Aluminium:
1 minute
Chromat:
3 minutes
Copper:
2 minutes
Iron:
1 minute
Hardness:
1 minute
Manganese:
5 minutes
Silicate:
5 minutes
3 % of scope of measuring range (all parameters)
15 ml / measurement (all parameters)
Aluminium:
3 x 0.2 ml / measurement
Chromat:
2 x 0.2 ml / measurement
Copper:
2 x 0.2 ml / measurement
Iron:
1 x 0.2 ml / measurement
Hardness:
2 x 0.2 ml / measurement
Manganese:
3 x 0.2 ml / measurement
Silicate:
3 x 0.2 ml / measurement
0 ... 72 h
0 ... 72 h
3 months
30 minutes / week
5 ... 40 °C
IP 43
StamoLys CA 70
Endress+Hauser