Installing The Measuring Sensors Prosonic Flow P/W (Clamp On) - Endress+Hauser PROline Prosonic Flow 93 FOUNDATION Fieldbus Operating Instructions Manual

Ultrasonic flow measuring system
Hide thumbs Also See for PROline Prosonic Flow 93 FOUNDATION Fieldbus:
Table of Contents

Advertisement

PROline Prosonic Flow 93 FOUNDATION Fieldbus
!
Endress+Hauser
3.3.4
Installing the measuring sensors Prosonic Flow P/W
(Clamp On)
1 or 3 traverses version
1.
Fix a tensioning band for small or large nominal diameters as described on
Page 17.
Install the second tensioning band (threaded stud on the opposite side). The
second tensioning band must still be moveable.
Intermediate step:
The sensor distance and the wire length are determined by means of the "Sensor" Quick
Setup (for devices with a local display). Run the Quick Setup as explained on Page 72.
The sensor distance is displayed in the "Sensor distance" function (6886) and the wire
length in the "Wire length" function, (6885).
Note!
• The transmitter must have a local display and be installed and connected to the power
supply before you can run the "Sensor" Quick Setup.
• In the case of transmitters without a local display, the value can be determined as
explained on Page 63.
2.
Enter the wire length on both halves of the wire.
Fig. 13:
Marking the determined wire length on the wire measurement equipment (SL = wire length)
3.
Push the cable lug and the fixer over the first threaded stud. Lead each wire along
one side of the pipe. Push the cable lug and the fixer over the second threaded
stud. Pull in the threaded stud with the tensioning band until both wires are the
same length. Fix the tensioning band. Loosen the Phillips screws of the fixing parts.
Remove the wires.
Fig. 14:
Use of the wire measuring equipment for positioning the threaded studs
SL
1
2
3
4
5
6
3 Installation
7
8
9
10
11
21