GE Senographe 2000 D Operator's Manual page 161

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Message
Msg.
on Column
Source
No.
Readout
S95
AWS
None
S96
AWS
None
S97
AWS
None
S98
AWS
None
S99
AWS
None
S100
AWS
None
Message on
Explanation
Control Console
S95 Medical
AWS has not yet
Application not
entered application
active
mode. This message
is always displayed
after system boot and
after each close exam
even during review.
In fact it is displayed
because no application
are ready to get the
S96 System
A diagnostic or
activity conflict
calibration application
runs while the medical
application is active.
S97 AWS reset in
The AWS is
progress, wait
completing a reset
S98 Exp.
The current application
inhibited CAL/IQ
is CALtool or IQtool.
in progress
No exposure is
requested by this
S99 OFFSET
The current application
ACQUISITION IN
is CALtool or IQtool.
PROGRESS
Offset acquisition has
S100 PLEASE
AWS is being
WAIT ...
reconfigured. The
message is displayed
until reconfiguration is
159
Senographe 2000 D Acquisition System
Start a new exam or QAP
acquisition.
image.
Check if several application is
running if yes close one.
No action required, wait for
completion of the reset.
operation.
No action required.
application.
No action required.
started.
No action required.
complete.
OM 5179217–1–100