GE Senographe 2000 D Operator's Manual page 37

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Note:
As a safety measure, the compression system is fitted with magnetic braking to avoid the
paddle falling in the event of a power cut. If such a power cut occurs during an examination, a
force of around 5 daN could remain on the compression paddle. To disengage the patient,
raise the paddle steadily (do not try to lift it quickly) to counteract this compression force.
Note:
Pressing a compression pedal or moving one of the thumbwheels illuminates the centering
light.
D The Magnification Platform
holes provided for the selected magnification (located on either side of the Receptor–arm).
To unlock the magnification platform, press the lever located on the left of the platform and pull the
platform forward (as shown on the label on top left side of platform). When the magnification
platform is inserted the small focal spot is automatically selected, and the FOV is reduced to 9 x 9.
Magnification factors are defined in the middle of a 4 cm thick breast, i.e., 2 cm above the
magnification platform. The useable magnification factors corresponding to each pair of holes, with
an SID of 660 mm, are:
Middle holes: 1.8
D
Bottom holes: 1.5
D
Do not use the upper holes for insertion of a magnification platform during
NOTICE
patient examinations
Always make sure that the magnification platform is correctly locked in place
NOTICE
before starting an exam.
D The Image Receptor
below).
6
is installed by latching the platform onto the Receptor–arm in the
7
functions as a breast support and includes a removable grid (see
35
Senographe 2000 D Acquisition System
OM 5179217–1–100