GE Senographe 2000 D Operator's Manual page 152

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Message
Msg.
on Column
Source
No.
Readout
I19
GEN
None
I21
GEN
None
I22
GEN
None
I23
GEN
None
I24
GEN
None
I25
GEN
None
I27
GEN
None
Message on
Explanation
Control Console
IDC/SENO
Image Detection
comm. lost
Control system cannot
communicate with
If the communication is
not recorver in few
minutes, the sysetm
will be in failure mode
Heating failure
Molybdenum focal
track 1
track is out of use.
Heating failure
Rhodium focal track is
track 2
Console
Generator-Control
communication
Console connection is
fail
Line power
Micro power cut
interruption
occurred on line
Recumbent
Operator has selectec
patient view
a view name for
names
recumbent patient.
Generator CPU
CPU backup battery is
battery error
discharged.
150
Senographe 2000 D Acquisition System
1. Wait that the Image Detector
control system finished to recorver
itself.
Generator.
(error E19)
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
You can continue using rhodium
focal track.
Switch OFF Gantry, then switch ON.
out of use.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
You can continue using molybdenum
focal track.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
faulty.
messages and contact GEMS Field
Service.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
supply.
messages and contact GEMS Field
Service.
If you don't want to run acquisition
with this selected view name, enter
again to the set–up/Proj. menu
without valid to come back to normal
view name selection.
Else the view name selection will
come–back to normal selection after
the next exposure.
Contact GEMS Field Service.
DO NOT REMOVE POWER FROM
EQUIPMENT.
OM 5179217–1–100