GE Senographe 2000 D Operator's Manual page 159

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Message
Msg.
on Column
Source
No.
Readout
S82
AWS
None
S83
AWS
None
S84
AWS
None
S85
AWS
None
S86
AWS
None
S87
AWS
None
Message on
Explanation
Control Console
S82 Image
One of the image
detection failure
detection coponent is
S83 Detector
The environment of
environment not
the detector is not
OK
correct to allow
S84 Image
The temperature drifts
quality is
slowly from the optimal
degraded
value specified for best
Image Quality.
S85 Detector
The environment of
environment
the detector is failed.
failure
S86 AWS/IDC
AWS system cannot
Communication
communicate with
lost
Image Detection
control system but the
communication should
be recorved itself.
S87 AWS/IDC
AWS to Image
Communication
Detection control
failure
system communication
157
Senographe 2000 D Acquisition System
1. Switch OFF the whole system and
wait for the end of the shutdown
failed.
sequence.
2. Restart the system.
3. If the problem persists, note ALL
messages and inform GEMS Field
Service.
1. Wait for completion of system
setting.
2. If this X–ray inhibition stays
acquisition.
displayed more that 10 mn
switch OFF the whole system and
wait for the end of the shutdown
sequence.
3. Restart the system.
4. If the problem persists, note ALL
messages and inform GEMS Field
Service.
Check image quality on current
exam.
If the problem persists, note ALL
messages and inform GEMS Field
Service.
1. Switch OFF the whole system and
wait for the end of the shutdown
sequence.
2. Restart the system.
3. If the problem persists, note ALL
messages and inform GEMS Field
Service.
1. Wait that the Image Detector
control system finished to recorver
itself.
1. Switch OFF the whole system and
wait for the end of the shutdown
sequence.
is failed.
2. Restart the system.
3. If the problem persists, note ALL
messages and inform GEMS Field
Service.
OM 5179217–1–100
Corrective Action