The selected items are copied from the CD-R to the workstation database.
D Click on the (Return to Browser) button to return to the main Browser.
D See section 2-4 on how to eject the CD from the CD-R drive after the restore operation.
Note:
During the restore, no other operations are possible on the Workstation.
When a patient name already exists in the Workstation database, any exams being restored
from the CD for that patient are appended to the existing exams.
Notes
2-3-2
The contents of the Media Browser: are up-to-date when the Media Browser: is first
opened. However, changes (such as after a save operation) are not displayed
automatically. Click on the (Refresh) button to update the Browser lists. The time of the
latest refresh is displayed on the status line.
The Media Browser: Tools menu contains the [Application management] menu item.
This item is for use of GE Service only.
The Main Browser and Media Browser: display the same columns of information in the Patients,
examinations, series and images lists. However, some of these columns may be empty in the Media
Browser:r. The full information will be displayed in the Main Browser after the images have been
restored.
Errors
2-3-3
If a problem occurs, an error message is displayed:
D The following message may be displayed during preparation for saving files and needs specific
action:
Save selection exceeds CD capacity. Save aborted
Explanation: You have selected more patient files than can be stored on a single CD.
Action: Select fewer patients to be saved on the CD.
If this message appears with only a single patient selected, you must delete some of the exams for
this patient before you can archive the patient.
D If a problem occurs during the save process, the error message Save failed is displayed.
First check that the CD-R drive is switched on, the blank recordable CD is correctly inserted in the
CD-R tray, and the tray is fully closed.
If the items above are correct, an error has occurred while recording the CD, and you will have to
repeat the save operation using a new blank CD.
D When restoring files, the operation is halted if there is not enough space on the Workstation disk to
restore the selected items (patients or exams). You can repeat the operation after freeing
Workstation disk space by deleting some patients that have already been saved.
Senographe 2000 D Acquisition System
135