GE Senographe 2000 D Operator's Manual page 149

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Message
Msg.
on Column
Source
No.
Readout
E51
GAN
Collim. Fail
E53
GAN
Collim. Fail
E54
GAN
Failure
E55
GAN
Housing Fail
E58
GAN
Collim. Fail
E64
GAN
Filter Fail
E71
GAN
Bucky Fail
E76
GAN
Ang. Failure
Message on
Explanation
Control Console
Lateral Collimator
A problem has been
Failure
reported in the lateral
collimator, so correct
format cannot be
Exposure is inhibited.
Rear Collimator
A problem has been
Failure
reported in the rear
collimator, so correct
format cannot be
Exposure is inhibited.
Gantry Electronic
A problem has been
Failure
reported by the Gantry.
Housing
A problem has been
Angulation
reported in tube
Failure
housing angular shift.
Front Collimator
A problem has been
Failure
reported in the Front
collimator blade, so
correct format cannot
Filter Positionning
A filter rotation
Failure
problem has been
reported, so good filter
positioning is not
Exposure is inhibited.
Bucky Failure
Bucky cannot be
moved under normal
Angle Sensor
Fault in Gantry angle
Failure
147
Senographe 2000 D Acquisition System
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
ensured.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
ensured.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
be ensured.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
ensured.
1. Remove Bucky.
2. Reinstall Bucky.
conditions.
3. Make a Rad Prep.
4. If failure continues, note ALL
messages, and contact GEMS Field
Service.
Exposures can be made without the
Bucky.
Switch OFF Gantry, then switch ON.
sensor.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
OM 5179217–1–100
Corrective Action

Advertisement

Table of Contents
loading

Table of Contents