GE Senographe 2000 D Operator's Manual page 36

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Receptor-arm Support
2-1-2
The main components of the Receptor-arm Support are listed below:
D Guide Rail
3
and Compression system including magnification platform and compression
paddle.
D The compression system comprises the paddle holder
knobs
2
, and the interchangeable compression paddle
Compression and decompression are controlled manually or by the foot pedals
the floor.
Maximum compression force, speed of compression and decompression height are
programmed by the system SET-UP functions.
Manual adjustments to compression force can be made using the knobs
side of the compression paddle arm. Turn the top of the knob towards the patient to increase
compression, away from the patioent to decrease compression.
To change a compression paddle: slide the paddle arm sideways from the compression paddle
support.
6
5
7
Senographe 2000 D Acquisition System
8
1
4
8
1
, the manual compression adjustment
4
.
34
OM 5179217–1–100
2
3
5
placed on
2
, located on each