GE Senographe 2000 D Operator's Manual page 170

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
2-4
Attenuation Equivalence
Attenuation equivalence for components in the X-ray beam (in accordance with FDA HHS 21 CFR,
§ 1020.30; measurements made at a potential of 100 kVp, using an X-ray beam with an HVL of
2.7 mm of aluminum).
Magnification device
Image receptor support
3.
AMBIENT CONDITIONS
3-1
Operational Ambient Conditions
D Humidity:
D Temperature
D Atmospheric pressure:
3-2
Transport and Storage Ambient Conditions
D Humidity:
D Temperature
D Atmospheric pressure:
3-3
Ambient Light Level
Senographe 2000 D monitors are adjusted for use in an optimum light level of 50 lux.
4.
TUBE INFORMATION
Tube Cooling Curves and other tube-specific technical parameters can be found in the Tube Product
Data document, part no. 2106037–100, included with the system documentation.
Component
Al equivalence (mm)
Bucky
10% min., 80% max.
15_C (59_F) min., 35_C (95_F) max.
700 hPa min., 1060 hPa max.
10% min., 50% max.
95% max. no longer than two weeks.
+10_C (50_F) min., 25_C (77_F) max.
–10_C (14_F),
50_C (122_F) no longer than 1 day.
500 hPa min., 1060 hPa max.
Senographe 2000 D Acquisition System
less than 0.2
less than 0.3
less than 0.3
168
OM 5179217–1–100