GE Senographe 2000 D Operator's Manual page 154

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Message
Msg.
on Column
Source
No.
Readout
I56
GAN
Slipping
I62
GAN
Oscillations
I70
GAN
Comp
Failure
I71
GAN
Light Fail
I72
GAN
Column Fail
I73
GAN
Rotat. Fail
I74
GAN
Lock Failure
I75
GAN
Button Fail
I76
GAN
Failure
I77
GAN
Pedal Fail
Message on
Explanation
Control Console
Compression
Compression motor
Motor Slipping
has slipped and lost its
Arm oscillations
Gantry oscillate then
the sensor angle value
is not stable.
This angle variation
could affect view name
Compression
Compression did not
Failure
function correctly.
Light Centering
The light centering
Device Fail
device is not
functioning correctly.
Column Motion
Elevator is not
Failure
functioning correctly.
Rotation Brake
Rotation brake is not
Failure
functioning correctly.
Arm Lock Failure
Arm lock is not
functioning correctly.
Button Security
Problem reported in
Failure
operator presence
interlock in control
Arm Sensor
An error has occurred
Failure
in the rotation angle
Comp Pedal
The operator presence
Security Failure
interlock for the
compression pedal did
not function correctly.
152
Senographe 2000 D Acquisition System
1. Release footpedal or compression
button, and move paddle up to park
position.
position.
2. Repeat the compression.
1. Wait that the gantry stabilizes.
calculation.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
buttons.
Service.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
sensor.
messages and contact GEMS Field
Service.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
OM 5179217–1–100
Corrective Action