GE Senographe 2000 D Operator's Manual page 54

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
6.
VIEW NAME SELECTION
D View names (Projection Position Codes) used by the Senographe 2000 D system are described
below in the section View Names.
D View names are normally selected automatically, according to the selected laterality, the presence
or absence of the magnification platform, and the gantry angle, as described in the section
Automatic View Name Selection.
D For views which do not correspond to the automatic selection, the operator must select additional
information manually, as described below in the sections Special Views Selection and View
Names for Recumbent Patients.
6-1
View Names
View names used by the Senographe system are shown here. They are based on the ACR (American
College of Radiologists) standardized abbreviations for mammography projection position codes.
Description
Prefixes selected by operator (laterality):
Right
Left
Prefix selected automatically (magnification):
Magnification
View Names selected automatically (applicable only to standing or sitting patients):
CranioCaudal
MedioLateral Oblique
MedioLateral (90_)
LateroMedial (90_)
LateroMedial Oblique
CranioCaudal (from below)
Superolateral to Inferiormedial Oblique
View Names selected manually:
Exaggerated CranioCaudal
Exaggerated CranioCaudal
Cleavage
Axillary Tail
Suffixes selected manually
Implant Displaced
Tangential
Spot Compression
Roll Lateral
Roll Medial
Prefixes and suffixes are used with the main view names; for example, RMMLO equals Right
Magnified MedioLateral Oblique, and LCCRL equals Left CranioCaudal with upper breast tissue
Rolled Laterally.
Senographe 2000 D Acquisition System
Code
R
Laterality selected by operator
L
Laterality selected by operator
M
Selected automatically if the Mag. stand is present
CC
Inferior side of breast against receptor
MLO
Inferior lateral side of breast against receptor
ML
Lateral side of breast against receptor
LM
Medial side of breast against receptor
LMO
Superior medal side of breast against receptor
FB
Superior side of breast against receptor
SIO
Inferior medial side of breast against receptor
XCCL
As CC with patient rotated laterally
XCCM
As CC with patient rotated medially
CV
As CC but inferior side of both breasts against receptor
AT
As MLO; axillary breast medially and anteriorly onto receptor
ID
Implant pushed back and flattened against chest wall
TAN
Area of interest projected close to skin surface
S
Spot compression with or without magnification
RL
Roll the breast slightly in lateral direction
RM
Roll the breast slightly in medial direction